Configuring rapid software — given imaging limited capsh-3 справочник пользователя — страница 182 из 250

PillCam Capsule Endoscopy

172

RAPID Settings

3.

If more than one physician uses this installation of RAPID, select 

Clear physician name when new 

video opens

. If selected, you can manually enter the physician name for each video when needed. 

4.

In Facility details

, type in the facility information (for example, address and telephone number).

5.

To add your facility’s logo to reports, click 

Change Logo

a.

Navigate to the logo file (jpeg image) you previously saved on your computer and select the 
file. 

b.

Click 

Open

 to set the file as the report logo.

c.

To clear the selected logo and prevent any logo from appearing on the report, click 

Clear Logo

.

6.

Click 

Apply

 to accept the new settings or click 

Cancel

 to close the 

Settings

 screen without making 

any changes. Clicking 

OK

 will accept the new settings and close the 

Settings

 screen.

Regional Settings Section

By default, the RAPID software is set to English and uses metric units of measure. You may change these 
settings. 

To change the language:

1.

From the Home screen, select 

Tools 

Settings

. The 

Settings

 screen appears.

2.

Under the 

General

 tab, in the 

Regional Settings 

section, open the 

Language

 list and select the 

desired language.

3.

Enable spell check

 is selected by default. Un-check to disable spell check.

4.

Enable GI Dictionary

 is selected by default. Un-check to disable the GI Dictionary.

5.

Open the 

Full name format

 list and select the desired name format.

6.

Click 

Apply

 to accept the new settings or click 

Cancel

 to close the 

Settings

 screen without making 

any changes. Clicking 

OK

 will accept the new settings and close the 

Settings

 screen.

To change the unit of measure:

1.

From the Home screen, select 

Tools 

Settings

. The 

Settings

 screen appears.

2.

Under the 

General

 tab, in the 

Regional Settings 

section, open the 

Measurement units

 list and select 

the desired measurement system.

֠

֠

֠

֠

Note
The Logo used in the report is copied to a RAPID directory. The original is 
not affected.

Configuring rapid software — given imaging limited capsh-3 справочник пользователя — страница 201 из 250

Configuring RAPID Software

Regimen Manager

191

Print in Pocket Format

When enabled, the post-capsule ingestion instructions will be printed in a convenient pocket format 
with fold lines for folding to an easy to carry size (quarter fold). The pocket format will show the 
Instructions guide on one side and important contact information on the other. 

,QVWUXFWLRQV’XULQJ<RXU3LOO&DP

3URFHGXUH

3DWLHQWQDPHBBBBBBBBBBBBBBBBBBBB
3DWLHQW,’BBBBBBBBBBBBBBBBBBBB
3URFHGXUHGDWHBBBBBBBBBBBBBBBBBBBB


BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

,16758&7,21*8,'(

,QJHVW>(17(5/2&$/%5$1’1$0(2)
PJ’203(5,’21(@ZLWKDJODVVRI
ZDWHU&RQWLQXHFRPSOHWHIDVWQRDGGLWLRQDO
GULQNLQJXQWLOQH[WDOHUW

‘LOXWH>(17(5/2&$/%5$1’1$0(2)
PO62′,803 263 $7(@LQDJODVVRI
ZDWHUDQGGULQNLW’ULQNDWOHDVWOLWHURI
ZDWHURYHUWKHQH[WKRXU<RXPDUHVXPH
GULQNLQJDOOFOHDUOLTXLGVIUHHO

‘LOXWH>(17(5/2&$/%5$1’1$0(2)
PO62′,803 263 $7(@LQDJODVVRI
ZDWHUDQGGULQNLW’ULQNDWOHDVWOLWHURI
ZDWHURYHUWKHQH[WKRXU<RXPDFRQWLQXH
GULQNLQJDOOFOHDUOLTXLGVIUHHO

,QVHUW>(17(5/2&$/%5$1’1$0(2)
PJ%,6$&2′</683326,725<@
DFFRUGLQJWRLQVWUXFWLRQVLQSDFNDJHLQVHUW

 <RXPDHDWDOLJKWPHDO

(QGRI3URFHGXUH3OHDVHUHPRYH
HTXLSPHQWIURPRXUERG

$»#»

‘() 
%
((((

Fold lines

Fold the bottom part of 
the printout first

Given imaging pillcam rapid 7 medical equipment setup & maintenance pdf view/download

Download Setup & maintenance of Given Imaging PillCam RAPID 7 Medical Equipment for Free or View it Online on All-Guides.com.

Installing rapid software — given imaging limited capsh-3 справочник пользователя — страница 173 из 250

Setup Requirements

163

Appendix A1

Installing RAPID Software

Setup Requirements 

Set up your office to accommodate the new PillCam Capsule Endoscopy System. Review the following 
workstation specifications:

Four electrical outlets are required to connect the following components: workstation computer, 
monitor, printer, and one cradle. Each additional cradle requires an additional outlet.

RAPID Installation

Before installing RAPID on your computer, it is recommended to be aware of the following 
prerequisites and other information: 

Before installation, it is highly recommended to format the hard disk in NTFS format, if not already 
done.

You need at least 1.5 GB free space on your computer to install RAPID.

For high resolution monitors, RAPID v8.3 offers the capability to resize and maximize the video 
view.

֠

֠

֠

֠

Note

You may use a Given approved power strip.

You may connect your monitor directly to the workstation using a DVI 
cable (the DVI-VGA adaptor is not required).

!

Caution
Do not connect any component of the PillCam Capsule Endoscopy System to 
the same outlet together with any appliance or device that has a high power 
requirement (refrigerators, generators, devices with motors, etc.). When 
setting up the system, make sure that the total power requirements for all of 
the devices connected to a single outlet or circuit do not exceed the rated limit 
for that circuit. If you are not sure of the rated limit, please consult your 
maintenance department or an electrician.

Do not use a KVM (Keyboard, Video, Mouse) Switch with the PillCam 
Capsule Endoscopy System.

֠

֠

֠

֠

Note
Extra space is needed for air circulation and cable connectors behind the 
workstation.